We propose a multi-patient inverse modeling framework for identifying effective calcium and citrate diffusion coefficients in hollow-fiber hemodialysis devices. The approach relies on a coupled forward model combining axisymmetric fluid dynamics with multi-species convection-reaction-diffusion, together with a derivative-free optimization strategy to estimate membrane transport parameters from outlet concentration measurements. To account for inter-patient variability, physiological input parameters are first generated from clinical data and complemented by a patient-specific hydraulic calibration step, ensuring physical consistency across the synthetic cohort. The inverse problem is formulated as a global least-squares minimization aggregating residuals over multiple patients. Numerical experiments on synthetic data demonstrate multi-patient identifiability of the diffusion coefficients in the exact-data setting. Robustness with respect to measurement noise is subsequently assessed by perturbing observable outputs at various noise levels, and sensitivity analyses are performed to quantify the influence of membrane transport parameters on model predictions. The methodology is then